The "Bio Wright™" range of products is recognised as leading the market in contamination control equipment, offering both product and personal protection.
 
 
 
Bio Wright™ 1 

Class 1 Microbiological Safety Cabinet 

All safety cabinets are independantly tested by the Bio Safety Biological Unit at C.A.M.R. Porton Down, exceeding BS 5726: Part 1 1992. 

An open fronted cabinet, through which the operator can carry out manipulations within the work area.  Room air flowing into the cabinet eliminates the escape of aerosols from the work area and is exhausted through a high efficiency HEPA filter.  Thus providing operator protection only. 

  • Fully welded 304L stainless steel construction.
  • Fully radius corners
  • Digital aor inflow meter
  • Digital hours run meter
  • Hinged toughened glass visor, supported on gas struts giving safe mechanical sealing.
  • Easily fitted night blind 100% seal every time.

  • Low airflow alarm.
 
    Compliance  
    Standard Size  
    Model Type  
    Weight  
    Power Supply  
    Lighting  
    Outlet  
    Control Panel  

    Visor  

    Night Door  
    Maintenance 

    Fully with BS5762 Part 1: 1992 C.E. Mark 
    1000mm Wide x 1280mm High x 600mm Deep.  Other sizes available. 
    Recirculating or ducted 
    70kg 
    220/240v 50Hz Single Phase 
    Minimum 500 lux at working height 
    Ducted to variable discharge options RH, LH or Vertical 
    Smooth wipe down membrane fascia incorporating audible and visual alarms for open visor and low airflow. 
    Monoform construction incorporating toughened glass screen and twin gas strut support system. 
    BioNite™ Night door complete with gasket,locators and quick type fasteners. 
    Side access filter removal(exhaust HEPA(s)).


 
Bio Wright™ II 

Class II Microbiological Safety Cabinet  

All safety cabinets are independantly tested by the Bio Safety Biological Unit at C.A.M.R. Porton Down, exceeding BS 5726: Part 1 1992. 

An open fronted cabinet offering both operator and product protection giving sterile conditions over the product handling area.  This is achieved by recirculating a flow of filtered clean air over the work area together with an inflow of air through the work environment. 

  • Fully welded construction. 
  • No use of silicon sealants on any joints 
  • Digital air inflow meter 
  • Digital hours run meter 
  • Hinged toughened glass visor, supported on gas struts giving safe mechanical sealing. 
  • Easily fitted night blind 100% seal every time. 
  • Low airflow alarms (both inflow and downflow). 

  • Easily maintained. 
 
    Compliance  
    Standard Size  
    Model Type  
    Weight  
    Power Supply  
    Lighting  
    Noise Level  
    Control Panel  

    Visor  

    Night Door  
    Maintenance  

    Services 

    Fully with BS5762 Part 1: 1992 C.E. Mark 
    1200mm Wide x 1390mm High x 680mm Deep.  Other sizes available. 
    Recirculating or ducted 
    190kg 
    220/240v 50Hz Single Phase 
    Minimum 500 lux at working height 
    59dB(a) at 0.3 meters from aperture. 
    Smooth wipe down membrane fascia incorporating audible and visual alarms for open visor and low airflow. 
    Monoform construction incorporating toughened glass screen and twin gas strut support system. 
    BioNite™ Night door complete with gasket,locators and quick type fasteners. 
    Filter change from front of cabinet with primary components unaffected.  Control chassis folds down from front face and self levels.  Tri-HEPA filtration system (Recirc. Model) 
    All service fittings outside contaminated airways. 

 
 
Bio Wright™ III  

Class III Microbiological Safety Cabinet  

All safety cabinets are independantly tested by the Bio Safety Biological Unit at C.A.M.R. Porton Down, exceeding BS 5726: Part 1 1992. 

This cabin separates the operator fron the product by a physical barrier i.e. it is a sealed enclosure with gloves mechanically attached to the visor.  An inlet HEPA filter allows clean air to be flushed through the system and exhausted via a HEPA filter to the outlet. 

  • Fully welded 304L stainless steel construction. 
  • Fully radius corners 
  • Digital hours run meter 
  • Magnehelic gauge (showing chamber pressure) 
  • Hinged toughened glass visor, supported on gas struts giving safe mechanical sealing. 
  • Easily maintained. 

  • Low airflow alarm. 
 
    Compliance  
    Standard Size  
    Model Type  
    Weight  
    Power Supply  
    Lighting  
    Outlet  
    Control Panel  

    Visor  

    Maintenance

 
    Fully with BS5296 Part 1: 1992 C.E. Mark 
    1000mm Wide x 1280mm High x 600mm Deep.  Other sizes available. 
    Recirculating or ducted 
    70kg 
    220/240v 50Hz Single Phase 
    Minimum 500 lux at working height 
    Ducted to variable discharge options RH, LH or Vertical 
    Smooth wipe down membrane fascia incorporating audible and visual alarms for open visor and low airflow. 
    Visor  Monoform construction incorporating toughened glass screen and twin gas strut support system. 
    Maintenance Side access filter removal(inlet and exhaust HEPA(s)).



 
 
Bio Wright™ HLF  

Horizontal Laminar Flow Cabinet  

This is a Clean Air Module giving unidirectional outflow of clean air to better than class E of BS 5295.  It may be used on a variety of applications where product protection is critical. 

  • Polyester painted mild steel construction 
  • Stainless steel work surface 
  • Mechanically clamped HEPA filter, removable from front 
  • Clear acrylic sides 
  • Minihelic pressure gauge (indicating soiling of HEPA filter) 
  • Easily maintained. 

  • Can be tailor made to suit individual requirements. 
 
    Construction  

    Compliance  
    Standard Size  
    Model Type  
    Weight  
    Power Supply  
    Lighting  
    Control Panel  
    Pre-Filter  
    HEPA Filter  
    Maintenance 

    Polyester painted mild steel, with stainless steel work surface and 6mm clear acrylic side screens. 
    Fully with BS5295 and carries C.E. Mark 
    12 sizes available. 
    Recirculating 
    Depending on size 
    220/240v 50Hz Single Phase 
    Minimum 500 lux at working height 
    Smooth wipe down fascia incorporating switches and minihelic pressure gauge. 
    Cartridge typeglass filter media, rated at 85% arrestance - Ashrae EU3. 
    Aluminium "Minipleat" type mechanically clamped against gasket seal. 
    Front access filter for all services. 

 
 
Bio Wright™ Fume Cupboard 

Fume Cupboards 
 

  • This is a bench or table mountd fume cupboard design, offering excellent face velocity performance and containment results when tested in accordance with BS 7258: 1994. Giving operator protection.
    • Various internal finishes. 
    • Stainless steel sill 
    • Safety control system. 
    • Easily maintained. 
     
      Construction  

      Model Type  
      Weight  
      Power Supply  
      Lighting  
      Interior finishes  

      Workbase  
      Plumbed service  

      Electrical Services 

      Double skin assembly constructed from epoxy powder coated mild steel.  Inner chamber rear baffles are removable for cleaning and maintenance. 
      Ducted 
      Depending on size 
      220/240v 50Hz Single Phase 
      Minimum 500 lux at working height 
      Epoxy coated mild steel, solid grade laminate, P.V.C. polypropylene, one piece stainless steel with radius corners or cast epoxy resin. 
      A variety of workbases are available depending on application. 
      Quick connection systems are employed using nylon flexible pipework and push fittings, pipework terminates at high or lowlevel depending on site requirements. 
      Electrical wiring is arranged as apre-wired internal ring circuit terminating in a high level junction box. 


     
     
    Bio Wright™ Validation  

    Validation and Maintenance Services  

    As part of the Atlas Clean Air package of Containment products, we also offer the facility of a full amintenance and validation service for new and existing clients, on a variety of different manufacturers equipment. 
     

    • Safety Cabinet Validation to BS 5726 
    • Laminar Flow Cabinet Validation to BS 5295 
    • Fume Cupoard Service/Validation to BS 7258 
    • Clean Room Validation 
    • Equipment repair and filter change carried out as part of Vaidation Visit. 


     
    If you have any questions or queries please do not hesitate to contact us.
    E-mail metrotec@iol.ie
    Products ButtonAgencies ButtonProfile ButtonNews Button
    Unit 3, Dun Laoi, Union Quay, Cork, Ireland.
    Tel. +353 21 319531  353 21 320301
    Fax. +353 21 314259  Mobile 086 2552688

     Link to Apollo  Last updated 04/03/98.  Best viewed with Download Netscape